The pressure produced by different devices

Flow rate(L/min) EEP EIP
HFT-300 H-80 M Gap of EEP HFT-300 H-80 M Gap of EIP
10 0.94(0.46) 1.07(0.57) −0.03(0.02) − 0.61(0.58) −0.55(6.32) − 0.01(0.31)
20 1.70(0.80) 1.93(0.85)* 0.20(0.55) −0.08(0.55) −0.06(0.54)# 0.11(0.53)
30 2.31(0.82) 2.87(1.34)* 0.37(0.59) 0.28(0.54) 0.64(0.67)# 0.17(0.47)
40 3.25(1.19) 4.39(1.88)* 0.66(1.01) 0.96(0.79) 1.54(1.07)# 0.48(0.78)
50 4.47(1.81) 5.71(22.78)* 1.07(1.22) 1.49(1.24) 2.25(1.30)# 0.72(0.74)
60 6.01(2.21) 7.40(3.21)* 1.16(1.09) 2.27(1.41) 3.12(1.45)# 0.62(1.30)

Values are median (IQR), * represents P < 0.05 compared to EEP of HFT-300 at the same flow rate, # represents P < 0.05 compared to EIP of HFT-300 at the same flow rate