Skip to main content
. 2023 Dec;387(3):328–336. doi: 10.1124/jpet.123.001828

TABLE 1.

Intra- and interday precision and accuracy of agmatine analytical method in 5% BSA

% BSA in PBS
Intraday Interday
Concentration (ng/ml) Measured Concentration (Mean ± S.D.; ng/ml) Precision
(% RSD)
Accuracy
(% Bias)
Measured Concentration (Mean ± S.D.; ng/ml) Precision
(% RSD)
Accuracy
(% Bias)
5 5.15 ± 0.40 7.81 3.08 4.78 ± 0.69 14.5 −4.48
8 8.26 ± 0.23 2.73 3.27 8.33 ± 1.02 12.2 4.11
80 78.3 ± 0.97 1.23 −2.18 79.3 ± 1.67 2.10 −0.83
800 785 ± 14.7 1.87 −1.84 789 ± 23.9 3.03 −1.33
1000 967 ± 15.6 1.62 −3.34 980 ± 33.9 3.46 −2.05