Dilution integrity of agmatine in 5% BSA

Concentration (ng/ml) % RSD % Bias
Dilution Factor Predicted Actual
Agmatine (plasma) 1:10 500 467 ± 6.37 1.37 −5.88
1:100 500 479 ± 22.3 4.64 −4.73