Skip to main content
. 2023 Nov 21;13:20397. doi: 10.1038/s41598-023-47582-y

Figure 5.

Figure 5

Integrated cartridge performance with custom PCR insert. (a) Reactions containing gDNA or extracted cellular DNA from SiHa (HPV16-positive), HeLa (HPV18-positive), and/or MS751 (HPV45-positive) cells at the indicated input concentrations were run in triplicate in the NATflow cartridges with the custom PCR insert. Highlighted images indicate invalid tests due to cartridge failures that were excluded from analysis. (b) Mean ± standard deviation of the signal generation at test lines, as determined by the AX-2X-S lateral flow reader of the replicates shown in (a). Circles represent individual replicates, and circles beneath the x-axis indicate expected 16 or 18/45 positivity. Signal above the positivity threshold (dashed line) is only seen at test lines when the target HPV genotype is present.