Skip to main content
. 2023 Nov 8;8(46):43538–43545. doi: 10.1021/acsomega.3c03977

Table 5. Comparison of Previously Reported Phenobarbital Potentiometric Electrodes.

sensory material electrode type slope (mV/decade) detection limit (M) linear range (M) pH range (pH) time response (s) ref.
β-cyclodextrin liquid polymeric electrode –59.1 3.5 × 10–6 5.0 × 10–6 to 1.0 × 10–2 9–11 25 (38)
γ-cyclodextrin liquid polymeric electrode –62.0 7.0 × 10–6 8.0 × 10–6 to 1.0 × 10–2 9–11 25
tetraoctylammonium/phenobarbaturate solid contact –59.0 not reported 2.0 × 10–4 to 1.0 × 10–1 9–11.5 15 (39)
Ionophore I Paper-based –64.0 ± 1.1 2.0 × 10–6 4.0 × 10–6-1.0 × 10–3 7.5–10.5 <5 This work
Ionophore II –63.6 ± 1.3 5.0 × 10–7 1.0 × 10–6-1.0 × 10–3
Ionophore I GC-Solid Contact –61.9 ± 0.9 5.3 × 10–8 1.0 × 10–7-1.0 × 10–3
Ionophore II –56.6 ± 0.3 3.1 × 10–8 6.0 × 10–8-1.0 × 10–3