Table 2.

mpMRI bladder imaging protocol for 1.5 and 3.0 Tesla (T).

Parameter T2w DW DCE
Field strength 1.5 T/3 T 1.5 T/3 T 1.5 T/3 T
Sequence * FSE SS-EPI FSPGR
Plane orientation Multiplanar Axial Axial
FOV (mm) 220–250 250–300 250–300
TR (ms) 4000–5000 4500–6000 3.5–4.5
TE (ms) 80–120 60–80 (minimum) 1.2–2.2
Acquisition matrix 256 × 192–256 128 × 128 256 × 192–214
Slice thickness/gap (mm) 3–4/0 3–4/0 3–4/0
Number of excitations 1–2 4–12 + 1
Flip angles (FAs) (degree) 90 90 15
b-values (s/mm2) NA 0 and 800–1000, up to 2000 optional NA

Note: Precontrast T1 mapping with multiple flip angles (30°, 15°, and 5°) is preferred; * fast spin echo (FSE), single-shot echo-planar imaging (SS-EPI), and fast spoiled gradient echo (FSPGR). + To improve SNR, more excitations are required at higher b-values