Table 2.

Accuracy and precision obtained from the analysis of human plasma quality control samples spiked with bosentan and its metabolite analysed by LC with enhanced product ion mode CID/UVPD and UVPD. LOQQC limit of quantification QC, LQC low QC, MQC medium QC, HQC high QC used for calibration range 5 to 500 ng/mL

Amount added (ng/mL) Bosentan (CID + UVPD) Desmethyl bosentan (UVPD) n
CV (%) Accuracy (%) CV (%) Accuracy (%)
LOQQC2 5 16.7 112 7.6 116 5
LQC2 15 11.1 108 7.5 95.8 5
MQC2 50 9.2 99.8 7.6 91.1 5
HQC2 500 5 112 4.8 121 5