Skip to main content
. 2023 Nov 30;18(11):e0295170. doi: 10.1371/journal.pone.0295170

Table 2. The diagnostic performance of EpiTuub® Fecal Rotavirus Antigen Rapid Test Kit, Ethiopia, 2022.

Test performance parameter Performance (%) 95% CI
Sensitivity 75.5 71.9–79.2
Specificity 98.2 97.1–99.3
Positive predictive value 89.9 87.3–92.4
Negative predictive value 95.0 93.1–96.8
Prevalence 17.5 14.3–20.7

CI = confidence interval