Table 5.

Details of instrument output discrepancies with statistical significance.

Biomarker Instruments Mean 1 (pg/mL) Mean 2 (pg/mL) Δ (pg/mL) % Difference* 95% CI (pg/mL) P-value, adjusted
ApoE 200 vs FM3D 819.5 796.1 22.4 2.8% 4.6 to 40.1 0.0107
MMP-9 MPX vs 200 166.7 175.1 −8.4 5.0% −14.0 to −2.9 0.0020
MPX vs FM3D 166.7 174.1 −7.4 4.4% −14.0 to −0.8 0.0261
SDC1 MPX vs 200 13986.0 13690.0 295.3 2.2% 91.9 to 498.8 0.0031
200 vs FM3D 13690.0 13991.0 −301.1 2.2% −494.2 to −108.1 0.0015
*

Differences calculated as the absolute value of the A divided by the lower of the two mean protein values x 100, with % differences rounded to nearest 0.1%.

P-values calculated by repeated measures ANOVA with Tukey post-test correction for multiple comparisons. All concentration values rounded to nearest 0.1 picogram.

Instrument Abbreviations: 200 = Luminex 200; FM3D = FlexMap 3D; MPX = MagPix.