Skip to main content
. 2023 Dec 1;24:70. doi: 10.1186/s40360-023-00711-9

Table 3.

Nebulization performance of the five nebulizers with varying IgG concentrations in ISO breathing pattern

IgG conc. (mg/mL) Nebulizer Residual volume (%) Nebulization time (min) Output rate (mL/min)
1 JN-PARIr 37.7 ± 1.1 5.6 ± 0.1 0.225 ± 0.00
JN-PARIb 30.9 ± 0.6 4.4 ± 0.1 0.315 ± 0.01
SMN-U150 14.1 ± 2.8 5.2 ± 0.7 0.334 ± 0.02
VMN-SM1 9.5 ± 2.3 6.7 ± 0.1 0.269 ± 0.01
VMN-SM3 8.7± 2.0 5.2 ± 0.1 0.351 ± 0.01
10 JN-PARIr 35.4 ± 4.1 5.8 ± 0.8 0.223 ± 0.02
JN-PARIb 35.6 ± 7.0 4.6 ± 0.2 0.282 ± 0.05
SMN-U150 15.7 ± 0.7 5.3 ± 0.2 0.317 ± 0.01
VMN-SM1 13.6 ± 3.0 7.1 ± 0.6 0.246 ± 0.02
VMN-SM3 15.1 ± 2.4 5.4 ± 0.1 0.314 ± 0.01*
20 JN-PARIr 34.6 ± 3.0 6.1 ± 0.9* 0.216 ± 0.02
JN-PARIb 32.2 ± 1.9 4.9 ± 0.8 0.275 ± 0.02
SMN-U150 16.2 ± 2.4 5.8 ± 0.8 0.290 ± 0.01
VMN-SM1 11.6 ± 1.7 7.4 ± 0.2* 0.239 ± 0.00*
VMN-SM3 13.1 ± 0.7 6.3 ± 0.1* 0.275 ± 0.01*
40 JN-PARIr 28.2 ± 4.0 7.2 ± 0.9* 0.202 ± 0.02
JN-PARIb 24.6 ± 1.8 5.9 ± 0.7* 0.258 ± 0.02
SMN-U150 10.4 ± 4.4 6.4 ± 0.7* 0.281 ± 0.03*
VMN-SM1 19.6 ± 1.6* 8.1 ± 0.9* 0.199 ± 0.01*
VMN-SM3 13.7 ± 5.8 7.6 ± 0.7* 0.229 ± 0.01*

All data are presented as the mean ± SD values. * p < 0.05 versus 1 mg/mL of IgG