TABLE V.

Acoustic Output for M-Mode Sequence

Code (bits) ISPPA0.3 (W/cm2) ISPTA0.3 (mW/cm2) PNP0.3 (MPa) MI TIS TIB TIC

1 1.11 0.69 0.12 0.07 0.04 0.01 0.16
13 1.24 5.32 0.16 0.10 0.28 0.07 1.21
25 1.11 10.18 0.17 0.10 0.53 0.13 2.32
39 1.11 15.36 0.17 0.10 0.81 0.20 3.51
65 1.06 25.32 0.17 0.10 1.33 0.34 5.78