Table 2.

Imaging protocols. The first 101 brains were scanned using the original imaging protocol. The revised protocol incorporating a new diffusion-weighted sequence as well as a higher resolution T2 TSE sequence was used for the next 59 brains

Original TR (ms) TE (ms) Flip Angle Resolution (mm3) Scan Time (min) Other parameters
SWI/QSM 39 6.72, 12.79, 21.29, 29.79 15° 0.5×0.5×0.5 61 6 averages
T1-Weighted 2200 3.25 13° 0.5×0.5×0.5 24 4 averages
T2 TSE 3750 20, 30, 50 120° 1.5×0.6×0.6 40 4 averages
Multi-Echo GRE 35 5, 10, 15, 20, 25, 30 20° 1.0×1.0×1.0 5
Revised TR (ms) TE (ms) Flip Angle Resolution (mm3) Scan Time (min) Other parameters
SWI/QSM 39 6.72, 12.79, 21.29, 29.79 15° 0.5×0.5×0.5 61 6 averages
T1-Weighted 2200 3.25 13° 0.5×0.5×0.5 24 4 averages
T2 TSE 9790 23, 34, 68 120° 0.8×0.6×0.6 293 14 averages
DW True FISP 29.77 23.89 35° 1.0×1.0×1.0 472 b-value 1 = 0s/mm2 b-value 2 = 18000s/mm2 44 diffusion directions 8 B0 images