Fig. 1.
(A) Scheme of the microFIA system to determine Fe(III). (a) Peristaltic pump for solutions, (b) six-way valve, (c) diluter, (d) mixer, (e) disperser, (f) detector, (g) 3-way solenoid valve, (h) peristaltic pump for cleaning, (i) analytical signal, the peak height is proportional to concentration. (B) Analytical signal and sequence of time diagram to determine Fe(III): injection time, start-up time, time at the maximum signal, and recovery time of the baseline. The total analysis time is 10.1 min