Fig. 3.
Calibration performed in the microFIA at optimal conditions. (A) Analytical signal (λ = 525 nm) at different concentrations, (B) blank and (C) calibration plot. The equation of the calibration curve is Voltage = 0.00063 (± 0.00002) [Fe(III)]—0.12 (± 0.09) (r2 = 0.996; n=5 in triplicate). Operational conditions are injection volume 25 µL, carrier flow rate 0.33 mL/min, diluent flow rate 0.39 mL/min, reagent flow rate 0.36 mL/min, and the ethanol flow rate 1.36 mL/min