Skip to main content
. 2023 Dec 2;191(1):3. doi: 10.1007/s00604-023-06029-x

Table 2.

Principal parameters of quality of the automatic microFIA system developed for determination of Fe(III)

Parameter Value
Sensitivity (V L mg−1) 0.00063 ± 0.00002
Detection limit (mg·L−1) 11
Quantification limit (mg·L−1) 25
Linear range (mg·L−1) 25–6000
Repeatability (CV) (n = 9) 0.69–4.92%
Reproducibility (RSD) (n = 7) 3.1%
Selectivity At pH acids not interference was observed
Sample throughput (h−1) 6
Accuracy (RV%) 99.9 ± 1.1