Skip to main content
. 2023 Dec 2;13:21301. doi: 10.1038/s41598-023-48082-9

Table 1.

Main material and LIPUS parameters used in the COMSOL simulation.

Domain Parameter Value Unit
LIPUS Transducer Frequency 1.0 MHz
Acoustic Power 8.4 W
Duty Cycle 50%
Transducer Diameter 24 mm
Water Attenuation Coefficient 0.0022 dB/(cm MHz2)
Density 994.23 kg/m3
Speed of Sound 1520.6 m/s
Absorption Coefficient 0.0253 Np/m
Acrylic Holder Density 1190 kg/m3
Young's Modulus 3.2 GPa
Poisson's Ratio 0.35
Nylon Clips Density 1150 kg/m3
Young's Modulus 2.0 GPa
Poisson's Ratio 0.4

Material parameters were taken from57,58.