Table 5.

CIED lead function measurements

n Pre-LAAO Post-LAAO Degree of change
Total n Mean ± SD Mean ± SD Mean ± SD p value
Right atrium
  Impedance (Ω) 19/22 461 ± 110 457 ± 106  − 4.4 ± 28 0.507
  Threshold (V) 14/22 0.85 ± 0.45 0.89 ± 0.52 0.05 ± 0.21 0.430
  Sensing (mV) 15/22 2.42 ± 1.47 3.15 ± 1.70 0.73 ± 1.43 0.069
Right ventricle
  Impedance (Ω) 26/31 528 ± 165 550 ± 215 21 ± 122 0.374
  Threshold (V) 26/31 1.08 ± 0.75 1.11 ± 0.72 0.03 ± 0.38 0.674
  Sensing (mV) 22/31 9.86 ± 4.56 10.47 ± 5.84 0.60 ± 4.98 0.576
Left ventricle
  Impedance (Ω) 4/5 472 ± 247 452 ± 221  − 20 ± 42 0.417
  Threshold (V) 3/5 1.42 ± 0.45 1.37 ± 0.32  − 0.05 ± 0.18 0.678
  Sensing (mV) 2/5 21.90 ± 2.69 19.8 ± 0.28 2.10 ± 2.40 0.433

CIED, cardiac implantable electronic device; LAAO, left atrial appendage occlusion; SD, standard deviation