. 2023 Dec 5;10:77. doi: 10.1186/s40658-023-00595-y

Table 3.

Measured data: transaxial and axial hPVC FWHM that resulted in the smoothed GT emission image that best matched the reconstructed image

Scanner Reconstruction hPVC(xy) FWHM [mm] hPVC(z) FWHM [mm]
D710 VPFX 2i (“EARL1”) 10.7 ± 0.06 7.2 ± 0.39
D710 VPFX 4i 9.6 ± 0.13 8.5 ± 0.25
D710 QCFX 8.9 ± 0.14 7.2 ± 0.18
Vision TOF 8.1 ± 0.11 6.8 ± 0.15
Vision TOF+PSF 6.7 ± 0.05 4.3 ± 0.18
Vision TOF (“EARL1”) 9.3 ± 0.07 8.3 ± 0.64

Uncertainty is one standard deviation of 100 voxel sub-sampling realisations, see “Appendix 3