. 2023 Oct 31;9(1):3–10. doi: 10.17996/anc.23-00005

Table 1. Methods for 123I-BMIPP imaging.

Imaging General procedures
Patient preparation Fasting at least 6 hours (6–12 hours) except for water intake
Scan Rest scan
Administered dose 74–148 MBq; intravenous injection
Time to imaging Early phase (20 min after injection); SPECT and planar imaging
Late phase (180–210 min after injection); SPECT and planar imaging
Imaging General parameters
Field of view Cardiac or chest
Image type Cardiac or chest SPECT and planar
Position Supine
Energy window 159 keV ± 10%
Matrix Planar, 256 × 256; SPECT, 128 × 128; at least 64 × 64
Pixel size 4.0–6.0 mm
Planar imaging Specific parameters
Views Anterior
Image duration 300 sec
Magnification × 1
SPECT imaging Specific parameters
Angular range Recommended, 180°; Optional, 360°
Detector configuration Recommended, 90°; Optional, 180°
ECG gating Non-gated image for WR calculation; 16 gates (optional 8 gates) for ECG gated acquisition
Viewangle 4.0°–6.0°, step and shoot
Number of views 30–120 sec (adjusted depending on projections)
Time per stop 30–60 sec (adjusted depending on projections)
Magnification × 1.5 (adjusted for scinticamera field of view and patients' stature)

ECG, electrocardiography; WR, washout rate.

Imaging parameters can differ among camera vendors, therefore appropriate adjustments can be made within those recommended above.