9, 33 |
DMSO conditioning droplet tests show residual DMSO on nozzle affecting droplet performance |
Residue on the surface of the nozzle is causing DMSO to persist. |
Dab the droplet of residual DMSO with a lint free wipe to dry the nozzle surface |
10,34 |
DMSO droplet images reveal dispensing inconsistency. |
Aberrant DMSO dispensing issue, occurs rarely. |
Adjust field files correspondingly, removing those with successful dispensing, and redispense reagent to relevant slide positions. |
17 |
Nozzle clogged during cell dispensing |
Large cell debris blocked the nozzle outlet and reduced droplet stability. |
Pause run, flush sample, run AirEx task, and aspirate new cell suspension to continue. |
25 |
Trypsin QC spot presence decreases in later Fields |
Undissolved protein or debris from Trypsin solution occluded the orifice during aspiration reducing the total volume aspirated |
Clear dispense fields with recognizable QC spot, restart run to dispense Trypsin to all arrays with poor trypsin QC spot |
26 |
Droplets may dry or swell overnight |
Improper setting for dew point correction factor |
The suggested setting is +2 C but this value may need to be slightly increased or decreased depending on lab. |
33 |
Label dispense failure |
Improper label pickup or occasional error. |
Load the field file for the label that failed and initiate redispense. Repeat for each failed label. |
39 |
TEAB dispense failure |
Aberrant TEAB dispensing issue, occurs rarely. |
Adjust field files correspondingly, removing those with successful dispensing, and redispense reagent to relevant slide positions. |
52 |
Residual reaction volumes after pickup are not negligible. |
Target teach height is too high from the slide surface to collect pooling volume completely |
Decrease the target teach height to ~1100 um from slide surface and restart the same pickup run to the same plate |