Skip to main content
. 2023 Nov 22;13(23):3507. doi: 10.3390/diagnostics13233507

Table 1.

Diagnostic parameters of STANDARD M10 and Xpert Xpress SARS-CoV-2 and Xpert Xpress Flu/RSV.

Virus/Parameters
% (95% CI)
Xpert Xpress SARS-CoV-2/Xpert Xpress Flu/RSV STANDARD M10
SARS-CoV-2
Sensitivity 100.0 (92.8–100.0) 100.0 (92.8–100.0)
Specificity 98.0 (89.2–100.0) 98.0 (89.2–100.0)
PPV 98.0 (89.4–100.0) 98.0 (89.4–100.0)
NPV 100.0 (92.6–100.0) 100.0 (92.6–100.0)
Accuracy 99.0 (94.5–100.0) 99.0 (94.5–100.0)
IAV
Sensitivity 97.9 (88.9–100.0) 97.9 (88.9–100.0)
Specificity 100.0 (93.8–100.0) 100.0 (93.8–100.0)
PPV 100.0 (92.5–100.0) 100.0 (92.5–100.0)
NPV 98.3 (90.9–100.0) 98.3 (90.9–100.0)
Accuracy 99.1 (94.9–100.0) 99.1 (94.9–100.0)
IBV
Sensitivity 100.0 (93.3–100.0) 98.1 (89.9–100.0)
Specificity 100.0 (93.8–100.0) 100.0 (93.8–100.0)
PPV 100.0 (93.3–100.0) 100.0 (93.2–100.0)
NPV 100.0 (93.8–100.0) 98.3 (90.9–100.0)
Accuracy 100.0 (96.7–100.0) 99.1 (95.1–100.0)
RSV
Sensitivity 100.0 (94.5–100.0) 100.0 (94.5–100.0)
Specificity 98.3 (90.8–100.0) 100.0 (93.8–100.0)
PPV 98.5 (91.8–100.0) 100.0 (94.5–100.0)
NPV 100.0 (93.7–100.0) 100.0 (93.8–100.0)
Accuracy 99.2 (95.6–100.0) 100.0 (97.1–100.0)