Table 1.
LumiraDx SARS-CoV-2 Ag Ultra (5 min) versus RT-PCR (Ct 30) |
LumiraDx SARS-CoV-2 Ag (12 min) versus RT-PCR (Ct 30) |
|
---|---|---|
Sensitivity | 100% | 100% |
Specificity | 98.4% | 99.4% |
Positive predictive value | 80.8% | 91.3% |
Negative predictive value | 100% | 100% |
Ct: cycle threshold; RT-PCR: reverse transcriptase polymerase chain reaction.