Table 3.

Parameters used for the operation of the mass spectrometer

Parameter Value
Polarity positive
Global settings
 Methods Duration (min) 160
 Ion Source Type NSI
 Spray Voltage: Positive Ion (V) 1800
 Ion Transfer Tube Temp (°C) 290
 Default Charge State 2
Full MS Scans
 Mass analyzer Orbitrap
 Automatic Gain Control (AGC) target 300%
 Microscans 1
 Orbitrap resolution 120,000
 Scan range (m/z) 350–1,800
 Max Injection Time (ms) 25
 AGC Target Custom
 DataType Profile
Filter Settings
 Include Charge State (s) 2–6
 Exclusion Duration (s) 35
 Mass tolerance (ppm) 10
Dependent MS/MS Scans
 Mass analyzer Orbitrap
 Automatic Gain Control (AGC) target 100,000
 Microscans 1
 Activation type HCD
 Normalized collision energy (%) 30
 Isolation Window 1.2
 First Mass 110
 Number of dependent scans 20 (from most intense precursors)
 Scan rate Normal
 Mass range Normal
 Charge exclusion Exclude +1 charge state
 Isolation Mode Quadrupole
 Ion trap scan rate Rapid
 Max Injection Time (ms) Custom
 Resolution 15000
 Maximum ion time 25 ms
 Loop count 10
 Isolation offset 0