Table VI.
Dilution Integrity at a Dilution Ratio of 1:1 for Spiked Whole Blood
Analyte | Intra-day precision (CV) | Inter-day precision (CV) | Accuracy (bias) |
---|---|---|---|
Metodesnitazene | 9 | 9 | 14 |
Flunitazene | 5 | 5 | 8 |
Metonitazene | 6 | 7 | 12 |
Etodesnitazene | 8 | 7 | 14 |
Etonitazepyne | 8 | 5 | 12 |
Clonitazene | 6 | 4 | 9 |
Etonitazene | 6 | 5 | 12 |
Isotonitazene | 6 | 4 | 10 |
Protonitazene | 5 | 7 | 12 |