Skip to main content
. 2023 Nov 15;11(6):e02078-23. doi: 10.1128/spectrum.02078-23

TABLE 3.

Statistical analysis of the sensitivity and specificity of Espline HTLV-I/II with a 95% confidence interval

Assay name Method Sensitivity (95% CI) Specificity (95% CI)
Serodia HTLV-I PA 100% (92.44–100) 100% (91.24–100)
Lumipulse HTLV-I/II CLEIA 100% (96.26–100) 100% (93.98–100)
Architect HTLV-I/II Score CLIA 99.19% (97.64–99.78) 99.40% (98.70–99.72)