Skip to main content
. Author manuscript; available in PMC: 2023 Dec 13.
Published in final edited form as: Toxicol Sci. 2018 Apr 1;162(2):509–534. doi: 10.1093/toxsci/kfx274

Table 2.

Comparison of Method Detection Limit (from OECD TG 456) and Reported LLOQs for HT-H295R

Hormone Family Steroid Hormone OECD TG 456
HT-H295R Assay
Method detection limit (ng/mL) Lower limit of quantitation (LLOQ) (ng/mL)  Upper limit of quantitation (ULOQ) (ng/mL)
Androgen Testosterone (T) 0.1 0.1 20
Dehydroepiandrosteronea NA 3 600
Androstenedione NA 1 200
Estrogen Estradiol (E2) 0.01 0.03 6
Estrone NA 0.03 6
Progestagen 17α-hydroxyprogesterone NA 0.2 40
17α-hydroxypregnenolone 5 1000
Progesterone 0.2 40
Pregnenolonea 2 400
Corticosteroid 11-Deoxycortisol 5 1000
Deoxycorticosterone 0.5 100
Cortisol 0.5 100
Corticosterone 0.5 100

The gray cells highlight the comparison of testosterone (T) and estradiol (E2), as these are the only 2 hormones with minimum method detection limits (MDL) in the performance criteria for the test guidelines. The LLOQs for all of the hormones measured in Karmaus et al. (2016) are listed.

a

Dehydroepiandrosterone and pregnenolone were excluded from further analysis in the work herein as 69.5% and 53.1% of the measured values for these 2 steroid hormones, respectively, were below the LLOQ.