Skip to main content
. 2023 Dec 1;10:1269833. doi: 10.3389/fcvm.2023.1269833

Table 2.

Valve function and major complications.

Edwards SAPIEN 3 Boston Lotus Edge
TCC COP P-value TCC COP P-value
AV max. PG (mmHg) 22.3 (20.8–23.8) 23.4 (21.9–25.0) 0.303 23.5 (21.8–25.2) 25.0 (23.2–26.9) 0.213
AV mean PG (mmHg) 12.0 (11.1–12.9) 12.7 (11.8–13.6) 0.275 12.7 (11.7–13.6) 14.1 (13.0–15.0) 0.056
Aortic regurgitation grade II/III 0% 0% n/a 0% 0% n/a
Device embolization 0% 0% n/a 0% 0% n/a
Coronary obstruction 0% 0% n/a 0% 0% n/a

Data are presented as percentages or mean ± SD.

AV mean/max PG, aortic valve mean/max pressure gradient; COP, cusp-overlap projection; TCC, three-cusp coplanar projection.