Table 1. Chromatographic and Spectrometric Parameters of the Analytes 1–3 and M1–M3.
| analyte | RT (min)a | Q1/Q3 (m/z)b | calibrated range (R2)c | LLOQ (nM)d | precision (%)e | accuracy (%)e |
|---|---|---|---|---|---|---|
| 1 | 5.02 (±0.01) | 319.1/275.2*, 273.2 | 15.6 – 500.0 (0.993) | 15.6 (S/N ≥ 31) | 2.7–17.7 | 96.3–104.7 |
| 2 | 4.38 (±0.01) | 481.2/118.9, 59.1* | 31.3 – 500.0 (0.990) | 31.3 (S/N ≥ 36) | 4.4–24.3 | 86.1–107.8 |
| 3 | 5.21 (±0.01) | 727.3/643.3*, 625.4 | 2.0 – 500.0 (0.998) | 2.0 (S/N ≥ 23) | 1.0–18.9 | 96.4–104.2 |
| M1 | 4.38 (±0.01) | 495.2/319.2*, 192.9, 174.8 | 2.0 – 250.0 (0.997) | 2.0 (S/N ≥ 39) | 2.0–16.1 | 83.7–109.5 |
| M2 | 4.88 (±0.01) | 495.2/319.2*, 174.9, 113.0, 84.9 | 0.5 – 62.5 (0.996) | 0.5 (S/N ≥ 18) | 2.6–24.7 | 85.2–106.5 |
| M3 | 5.09 (±0.01) | 495.2/319.1, 192.9*, 113.0, 59.0 | 1.0 – 62.5 (0.994) | 1.0 (S/N ≥ 15) | 2.3–18.1 | 88.6–108.7 |
| 4 (IS) | 5.46 (±0.01) | 315.1/271.2, 253.2* |
Calculated from plasma analyses and plasma calibration standards.
Quantifier is marked w/asterisk.
Calibrated range with precision <25% at LLOQ and accuracy 80–120%.
LLOQ: lower limit of quantitation defined as the smallest concentration in the calibrated range.
Values from back-calculated calibration standards in the plasma matrix (cf. Supporting Information).