Table 3.
Distribution of the conclusive measurements collected by the EmbracePlus wristband and performance of the SpO2 algorithm in terms of Arms, bias and MAE for different SaO2 ranges.
| Range of the reference SaO2 | Number of paired samples [n] | Number of paired samples [%] | Arms [%] | Bias [%] | MAE [%] | Upper 95% LoA [%] | Lower 95% LoA [%] |
|---|---|---|---|---|---|---|---|
| [67 100]% | 373 | 100.0 | 2.4 | 0.05 | 1.82 | 4.8 | −4.7 |
| [67 80)% | 93 | 24.9 | 3.2 | 1.6 | 2.59 | 7.3 | −4.2 |
| [80 90)% | 138 | 37.0 | 1.9 | 0.3 | 1.43 | 4.0 | −3.4 |
| [90 100]% | 142 | 38.1 | 2.2 | −1.2 | 1.68 | 2.5 | −4.9 |