Table 2.
Summary of colorimetric sensing of breath biomarkers.
| Breath biomarkers | Clinical correlation | Related diseases | Sensing probes | Substrates | Humidity concerns |
|---|---|---|---|---|---|
| Oxygen (O2) | Oxygen consumption related to metabolism. | Oxygen uptake, metabolic rate.95 | Lignin95 | Paper95 | The sensor is capable of selectively detecting oxygen in human breath at high humidity (100%).95 |
| Carbone dioxide (CO2) | CO2 production related to metabolism. | COPD (chronic obstructive pulmonary disease), asthma,96 and capnography.46, 97 | pH indicators, such as mcresol purple,98 cresol red,98 phenol red,46 and thymol blue.96 | Hydrophobic plastic film (ethyl cellulose) incorporated with a phase-transfer agent (tetraoctyl ammonium hydroxide)46 | The colorimetric CO2 sensor showed stable performance at 37°C and 100% humidity.46 |
| Acetone | Breath acetone is correlated with blood glucose (BG) and blood ketone levels.99, 100 | Diabetes99, 100 | Resorcinol35, 101 | Nafion membrane35, 101 | The incorporated organic acids played an important role in mitigating the humidity interference to the Nafion membrane catalysts in humid breath environment.35 |
| Nitric oxide (NO) | NO level is elevated in asthma with inflammatory. | Asthma102, 103 | Cobalt tetraphenylpor phyrin (CoTPP)104 and phenylenediam ine derivatives.105 | Polystyrene (PSt), ethylcellulose (EC), polycyclohexyl methacrylate (PCHMA),104 and microporous cellulose membrane.105 | The hydrophobic polymer matrix significantly improved the humidity resistance.104 A Nafion tubing was integrated to avoid the humidity interference.105 |
| Hydrogen sulfide (H2S) | An endogenous gaseous signaling molecule and potential therapeutic agent.106 | Bad breath107 | Lead(II) acetate,10 silver nanoparticles,108 and copper(II) complex.107 | Polyacrylonitrile nanofibers,10 nylon membrane,108 and Prozorb substrate.107 | Sensor can detect sub-ppm H2S gas in high relative humidity (~90% RH).10 The sensitivity of the sensor is retained in 50–96% RH, but decreased at low humidity level, e.g. 20% RH.108 |
| Ammonia (NH3) | Associated with blood urea nitrogen | Renal failure109–111 | pH indicators109,110 | Hydrogel film (with the combination of polyvinyl alcohol, titanium oxide nanoparticles, and hydroquinone)109 | The hydrogel film with the combination of other chemicals provides a versatile matrix to minimize the environmental interference.109 |
| VOCs | Volatile organic compounds (VOCs) related to metabolic changes within cancer cells. | Lung cancer8,91, 92 | Metalloporphyr ins, pH indicators, and dyes with large permanent dipoles.8, 91, 92 | Poly(vinylidene fluoride) membrane92 | Sensor is stable against humidity because of the high hydrophobicity of polyvinylidene fluoride.92 |