. 2005 Mar 30;33(6):e56. doi: 10.1093/nar/gni054

Figure 6.

Figure 6

Workflow of operational procedure for RNA quality assessment. Integrity of the RNA, once extracted and purified from cell lines, clinical or biological tissues samples, is controlled from the widely used bioanalyzer electrophoretic traces. As standard part of the Agilent analysis software (25), a RIN metrics is first calculated, scoring each RNA sample into 10 numerically predefined categories of integrity (RIN, from 1 to 10; N is a threshold value). As an independent control, a degradation factor metrics (DegFact, from 1 to ∞; N′ is a threshold value) may optionally be allocated to each RNA sample using the bioanalyzer-independent degradometer software (24). In a standard operating procedure, RIN and/or DegFact metrics will first be used as a standard exchange language to document RNA integrity and degradation, second to classify the RNA in homogeneous groups, and finally to select samples of comparable RNA integrity to improve the scheme of meaningful downstream experiments. The standard operating procedure will benefit from feedback information that will help users to define threshold integrity metrics values based on the results of RNA-based analyses.