Parameters required for system suitability tests of CZE method.

PHE PAP ACT PAR CAC PNP Reference values [17,18]
Migration time (min) 2.446 3.196 3.574 3.880 4.152 5.284 ---------
Resolution (Rs) 11.538 5.617 3.908 3.335 16.526 Rs< 1.5
Tailing factor (T) 1.10 1.15 1.13 0.98 1.03 1.08 T = 0.8–1.2
Selectivity factor (α) 1.31 1.12 1.09 1.07 1.27 α < 1
Column efficiency (N) 25836.43 38424.36 35302.76 31855.11 47644.66 105031.30 Increase with efficiency of the separation