Skip to main content
. 2023 Dec 22;22(1):e12556. doi: 10.1002/rmb2.12556

TABLE 1.

Comparison of current immunoassays for measuring the AMH.

Description MenoCheck® picoAMH ELISA Ultrasensitive AMH/MIS ELISA Gen II AMH ELISA Elecsys® AMH immunoassay Access AMH
Manufacturer Ansh Laboratories (TX, USA) Ansh Laboratories (TX, USA) Beckman Coulter Diagnostics (TX, USA) Roche Diagnostics International Ltd. (IN, USA) Beckman Coulter Diagnostics (TX, USA)
Test format Sandwich‐type immunoassay Sandwich‐type ELISA ELISA (2 site manual immunoassay) Automated immunoassay Automated immunoassay
Limit of detection 1.3 pg/mL 0.023 ng/mL 0.05 ng/mL 0.010 ng/mL ≤0.02 ng/mL
Limit of quantification (LoQ) with <20% CV 3.2 pg/mL 0.06 ng/mL 0.13 ng/mL 0.030 ng/mL ≤0.08 ng/mL

Abbreviation: CV, coefficient of variation.