Table 2.

Reference irradiation conditions for low energy x-rays.

Influence quantity Reference characteristics or value
Phantom material Water-equivalent plastic or PMMA
Chamber type Plane-parallel for low-energy x-rays
Reference point of chamber At the centre of the outside surface of the chamber window (or additional foil, if this is used for beams above 50 kV)
Measurement depth Phantom surface
SSD 30 cm
Field size at SSD 3 cm diameter or the closest field size to this that will be used clinically
Reference air pressure 1013.25 hPa
Reference air temperature 20 °C
Humidity 50%