Table 2.
Reference irradiation conditions for low energy x-rays.
| Influence quantity | Reference characteristics or value | 
|---|---|
| Phantom material | Water-equivalent plastic or PMMA | 
| Chamber type | Plane-parallel for low-energy x-rays | 
| Reference point of chamber | At the centre of the outside surface of the chamber window (or additional foil, if this is used for beams above 50 kV) | 
| Measurement depth | Phantom surface | 
| SSD | 30 cm | 
| Field size at SSD | 3 cm diameter or the closest field size to this that will be used clinically | 
| Reference air pressure | 1013.25 hPa | 
| Reference air temperature | 20 °C | 
| Humidity | 50% |