Characteristics of some recommended cylindrical ionisation chambers for reference dosimetry in medium energy x-rays.

Chamber type Volume (cm3) Wall material Central electrode Wall thickness (mg cm-2) Waterproof? Manufacturer
30010 0.6 graphite & PMMA aluminium 56.5 no PTW1
30012 0.6 graphite aluminium 79 no PTW1
30013 0.6 graphite & PMMA aluminium 56.5 yes PTW1
30015 / 23331 1.0 graphite & PMMA aluminium 73 no PTW1
30016 / 23332 0.3 graphite & PMMA aluminium 67 no PTW1
31013 0.125 graphite & PMMA aluminium 78 yes PTW1
2561 0.325 graphite aluminium 90 no NE2
2611 0.325 graphite aluminium 90 no NE2
2571 0.6 graphite aluminium 65 no NE2
1

PTW: PTW-Freiburg GmbH, Freiburg, Germany.

2