Table 1.

Contrast and amplitude (voltage/threshold) generated by 100 μm, 40 μm and 20 μm pixels for the smallest resolvable C (gap width = 1.2× pixel size) under 3 mW mm−2 irradiance, with 9.8 ms pulses at 30 Hz repetition rate. Stimulation thresholds are 11.7 mV and 19.7 mV in the best and the worst cases, respectively.

100 μm PRIMA
40 μm pixels
20 μm pixels
Bipolar
Monopolar
Contrast
(%)
Voltage/
Threshold
Contrast
(%)
Voltage/
Threshold
Contrast
(%)
Voltage/
Threshold
Flat, worst Basic 94 1.6 38 2.2 36 0.83
Pre-charged 92 1.3 75 0.45
Flat, best Basic 100 5.3 48 5.6 44 2.2
Pre-charged 100 3.7 87 1.3
Pillars, worst Basic 75 4.3 68 1.6
Pre-charged 100 3.6 100 1.2
Pillars, best Basic 75 7.9 68 3.0
Pre-charged 100 6.8 100 2.2