Table 3.

Performance characteristics for detecting positive emission tomography (PET) status in the Alzheimer’s Disease Research Center (ADRC) cohort at 0.160 (maximum of Youden’s J index) and 0.170 (likely PET-negative) cutpoints, and the effect of prevalence and bias (bold=performance based on measured prevalence of PET positivity in full ADRC cohort).

Plasma Aβ42/40 Cutpoint Prevalence (%) Bias (%) Sensitivity (95%, CI) Specificity (95%, CI) Accuracy (95%, CI) PPV (95%, CI) NPV (95%, CI)

0.160 40 none 0.911 (0.842–0.970) 0.758 (0.678–0.832) 0.820 (0.772–0.864) 0.719 (0.656–0.788) 0.926 (0.876–0.974)
0.160 34* none 0.857 (0.738–0.976) 0.659 (0.537–0.768) 0.726 (0.645–0.798) 0.563 (0.478–0.661) 0.900 (0.820–0.980)
0.160 34* 22 0.333 (0.167–0.500) 0.854 (0.756–0.939) 0.677 (0.605–0.742) 0.538 (0.346–0.737) 0.714 (0.663–0.771)
0.160 34* 2 0.762 (0.619–0.905) 0.671 (0.549–0.780) 0.702 (0.621–0.782) 0.542 (0.446–0.649) 0.846 (0.762–0.932)
0.170 40 none 0.990 (0.960–1.000) 0.537 (0.443–0.624) 0.720 (0.672–0.768) 0.592 (0.546–0.643) 0.988 (0.953–1.000)
0.170 34* none 0.976 (0.905–1.000) 0.476 (0.354–0.598) 0.645 (0.573–0.718) 0.488 (0.433–0.554) 0.975 (0.907–1.000)
0.170 34* 22 0.429 (0.262–0.595) 0.829 (0.732–0.915) 0.694 (0.621–0.766) 0.563 (0.400–0.732) 0.739 (0.681–0.802)
0.170 34* 2 0.929 (0.833–1.000) 0.537 (0.415–0.659) 0.669 (0.589–0.742) 0.506 (0.441–0.583) 0.936 (0.851–1.000)
*

Amyloid positivity by PET in the ADRC cohort—MCI group only

+10% bias in Aβ42 and -10% bias in Aβ40 = 22% total bias per Rabe, et al.13

+2% total measured bias in the LC-MS/MS assay