Table 2.
LoQs (nmol/L) derived from in-house testing vs manufacturer’s claims.
Steroid | LoQ (in-house) | LoQ (manufacturer) |
---|---|---|
Aldosterone | 0.078 | 0.039 |
Cortisol | 3.380 | 4.190 |
Cortisone | 0.283 | 0.411 |
Corticosterone | 0.231 | 0.505 |
11-Deoxycortisol | 0.416 | 0.087 |
21-Deoxycortisol | 0.121 | 0.078 |
DHEAS | 39 | 66 |
Estradiol | 0.015 | 0.092 |
Testosterone | 0.035 | 0.017 |
Androstenedione | 0.112 | 0.080 |
DHEA | 0.738 | 0.794 |
11-Deoxycorticosterone | 0.024 | 0.070 |
Dihydrotestosterone | 0.279 | 0.145 |
17-OHP | 0.045 | 0.121 |
Progesterone | 0.111 | 0.095 |
In-house testing: Sciex QTrap 6500+ mass spectrometer; manufacturer: Sciex QTrap 5500 mass spectrometer.
17-OHP, 17-hydroxyprogesterone; DHEA, dehydroepiandrosterone; DHEAS, dehydroepiandrosterone sulfate; LoQ, limit of quantification.