. 2023 Nov 20;4(12):765–776. doi: 10.1016/j.hroo.2023.11.014

Table 2.

Procedure details

Overall (N = 925) LLL (n = 655) SDL (n = 270) P value
Implanted device type <.001
 Single-chamber PM 75 (8) 53 (8) 22 (8)
 Dual-chamber PM 709 (77) 479 (73) 230 (85)
 CRT device 141 (15) 123 (19) 18 (7)
LBBAP lead implant time (min) 23 ± 11 24 ± 23 21 ± 15 .006
LBBAP implant fluoroscopy (min) 11 ± 9 12 ± 9 10 ± 8 .010
No. of screw attempts 2.6 ± 1.8 2.6 ± 1.8 2.5 ± 1.8 .644
No. of lead turns >15 308 (51)a 283 (63)a 25 (16)a <.001
LBBAP success 852 (92) 624 (95) 228 (84) <.001
LBBAP criteria <.0001
 LBBP criteria 575 (62) 427 (65) 148 (55)
 LVSP criteria 277 (30) 197 (30) 80 (30)
"r" prime in lead V1 810 (88) 588 (90) 222 (82) .009
LB potential 401 (43) 311 (47) 90 (33) <.0001
QRS transition during pacing threshold 385 (42) 278 (42) 107 (40) .508
V6 RWPT (ms) 82 ± 16 82 ± 16 81 ± 15 .365
V1–V6 interpeak (ms) 40±13 40±13 39 ±13 .379
Paced QRS axis <.001
 Inferior 311 (34) 246 (38) 65 (24)
 Intermediate 322 (35) 227 (35) 95 (35)
 Superior 264 (29) 171 (26) 93 (34)
Septal lead position <.001
 Basal 108 (12) 83 (13) 25 (9)
 Mid-basal 699 (76) 511 (78) 188 (69)
 Mid-apical 70 (8) 32 (5) 38 (14)
R-wave sensing (mV) 10.8 ± 5.7 10.9 ± 5.8 10.3 ± 5.3 .112
Pacing threshold (V) 0.86 ± 0.39 0.80 ± 0.34 1.0 ± 0.46 <.001
Impedance (Ω) 912 ± 34 948 ± 235 822 ± 208 <.001
Final paced QRS duration (from pacing spike) (ms) 155 ± 19 156 ± 20 153 ± 17 .016
Final paced QRS duration (only QRS) (ms) 115 ± 15 115 ± 15 114 ± 15 .193

Values are given as n (%) or mean ± SD unless other indicated.

LB = left bundle; LBBAP = left bundle branch area pacing; LBBP = left bundle branch pacing; LVSP = left ventricular septal pacing; PM = pacemaker; RWPT = R-wave peak time; other abbreviations as in Table 1.

Number of lead turns was evaluated in 610 patients (450 with LLLs and 160 with SDLs).