Table 3.

Electrical parameters in the different subgroups

Overall
Patients without structural heart disease
Last 100 patients
LLL (n = 655) SDL (n = 270) LLL (n = 280) SDL (n = 151) LLL (n = 100) SDL (n = 100)
Implant electrical parameters
 Sensed R-wave amplitude (mV) 10.9 ± 5.8 10.3 ± 5.3 11.2 ± 5.5 10.4 ± 5.4 9.9 ± 5.4 10.5 ± 5.4
 Unipolar pacing threshold (V) 0.80 ± 0.34 1.0 ± 0.46 0.79 ± 0.34 0.96 ± 0.41 0.96 ± 0.42 1.0 ± 0.55
 Unipolar lead impedance (Ω) 948 ± 235 822 ± 208 940 ± 239 788 ± 201 1041 ± 202 838 ± 211
 Paced QRS duration (from pacing spike) (ms) 156 ± 20 153 ± 17 149 ± 17 149 ± 15 164 ± 20 154 ± 15
 Paced QRS duration (only QRS) (ms) 115 ± 15 114 ± 15 111 ± 13 111 ± 14 121 ± 14 115 ± 14
Last follow-up electrical parameters
 Sensed R-wave amplitude (mV) 15.4 ± 7.2 13.4 ± 5.4 15.9 ± 7.5 12.9 ± 5.1 17.1 ± 8.5 13.1 ± 5.3
 Unipolar pacing threshold (V) 0.84 ± 0.94 0.88 ± 0.34 0.88 ± 0.87 0.95 ± 0.37 0.85 ± 0.36 1.01 ± 0.36
 Unipolar lead impedance (Ω) 481 ± 102 508 ± 123 505 ± 101 534 ± 102 512 ± 74 556 ± 89

Values are given as mean ± SD.

Abbreviations as in Table 1.

Indicates statistical significance (P <.05) in comparison of LLLs vs SDLs.

Median follow-up 16.4 [9.6–24.6] months.