Skip to main content
. 2023 Nov 13;54(1):82–95. doi: 10.1007/s00247-023-05789-1

Table 3.

The test–retest precision of Deeplasia

Rating method Precision (months)
Full dataset (n=149) With clinical BA (n=106)
Deeplasia 2.7 (2.5, 3.1) 2.4 (2.1, 2.8)
Clinical rating - 2.6 (2.3, 3.0)
Blinded rating 5.6 (4.9, 6.2) 5.8 (5.1, 6.7)

The clinical and a blinded manual bone age rating is estimated on patients with genetically-confirmed disorders (i.e. excluding intrauterine growth retardation and  non-diagnosed cases) within the German Dysplastic Bone Databse (GDBD). n refers to the number of images within the GDBD for which the interpolation residuals could be estimated. The precision is stated with a 95% confidence interval. GDBD German Dysplastic Bone Dataset