Skip to main content
. 2024 Jan 2;14(1):698–710. doi: 10.21037/qims-23-843

Table 1. Paired t-tests of SNR and CNR between CI/VMI groups and CI/VMI + O-MAR groups.

SNR and CNR CI VMI
50 keV 70keV 90 keV 110 keV 130 keV 150 keV
SNR
   CI 2.44±0.87 2.29±1.00 2.42±0.81 2.55±0.90 2.6±0.96 2.63±1.00 2.64±1.03
   O-MAR 2.82±1.40 2.39±1.12 2.68±1.17 2.88±1.26 3.03±1.39 3.11±1.47 3.15±1.51
   P value 0.003 0.541 0.022 0.008 <0.001 <0.001 <0.001
CNR
   CI 4.33±1.46 3.87±1.68 4.19±1.42 4.41±1.66 4.63±1.62 4.72±1.67 4.76±1.70
   O-MAR 6.16±2.20 5.18±2.22 5.5±1.98 5.76±2.03 5.95±2.18 6.03±2.26 6.06±2.29
   P value <0.001 <0.001 <0.001 <0.001 <0.001 <0.001 <0.001

Data represented as mean ± standard deviations. SNR, signal-to-noise ratio; CNR, contrast-to-noise ratio; CI, conventional image; VMI, virtual monochromatic image; O-MAR, orthopedic metal artifact reduction algorithms.