Table 2. Optimization of the HPLC Method for the Bioanalysis of 5-FU Using CCD.

factors and their levels
responses goal
design run A B
1 0.0 –1.0 peak retention time (min) <5
2 –1.0 1.0    
3 0.0 0.0 peak area (mAU) >10,000
4 0.0 0.0    
5 1.0 0.0    
6 –1.0 –1.0    
7 1.0 –1.0    
8 –1.0 0.0    
9 0.0 1.0    
10 0.0 0.0    
11 1.0 1.0    
12 0.0 1.0    
13 0.0 0.0    
statistical analysis
evaluation parameter low (−1) middle (0) high (+1) model p-value R2
A: mobile phase ratio 92:8 94.6 96:4 quadratic 0.0023 0.9784**
B: column temperature (°C) 30* 35 40** quadratic 0.0019 0.9597*

* and ** indicated the lower and the higher values, respectively.