Published in final edited form as: Proc SPIE Int Soc Opt Eng. 2018 Jul 6;10718:107180R. doi: 10.1117/12.2318029

Table 2.

Summary statistics for phantom texture in the synthetic 2D mammogram assuming lattice parameters w = d = 6.3 mm.

SYNTHETIC 2D MAMMOGRAM
Groups of Features Individual Features Mean (Std Dev) of 16 Auto-Timed Images Coefficient of Variation Percentile Rank of Mean Relative to Clinical Distribution
Group 1: Grey-level histogram Max 2.98 (0.0241) 0.00811 92.1%
Min −1.26 (0.0238) 0.0188 65.8%
Mean 0.0371 (0.00588) 0.158 * 55.4%
Sum 1.29 × 102 (20.5) 0.158 * 51.4%
Entropy 5.96 (0.0266) 0.00446 1.40% **
Kurtosis 5.98 (0.175) 0.0292 99.0% **
Sigma 0.730 (0.00917) 0.0126 98.7% **
Skewness 1.35 (0.0438) 0.0324 99.7% **
5th Percentile −0.764 (0.0141) 0.0184 71.6%
5th Mean −0.883 (0.0168) 0.0191 72.4%
95th Percentile 1.59 (0.0201) 0.0127 >99.9% **
95th Mean 2.02 (0.0190) 0.00941 99.8% **
Group 2: Co-occurrence Cluster shade 1.99 × 104 (4.69 × 102) 0.0236 >99.9% **
Energy 1.28 × 10−3 (9.95 × 10−5) 0.0778 * 99.6% **
Entropy 8.78 (0.0337) 0.00384 <0.1% **
Inertia 4.92 × 102 (6.35) 0.0129 >99.9% **
Correlation 9.28 × 10−4 (2.46 × 10−5) 0.0265 30.1%
Haralick correlation 9.49 × 106 (5.00 × 105) 0.0528 * 1.50% **
Inverse difference moment 0.107 (0.00324) 0.0302 99.9% **
Group 3: Run length Grey-level non-uniformity 72.0 (1.65) 0.0229 89.8%
Run-length non-uniformity 3.25 × 103 (10.4) 0.00320 <0.1% **
Run percentage 0.934 (0.00295) 0.00316 0.5% **
High grey-level run emphasis 2.26 × 103 (40.7) 0.0180 11.5%
Long run emphasis 1.00 (6.79 × 10−5) 6.79 × 10−5 94.5%
Low grey-level run emphasis 2.94 × 10−3 (1.26 × 10−4) 0.0429 99.4% **
Short run emphasis 1.00 (1.70 × 10−5) 1.70 × 10−5 3.40%
*

Coefficient of variation exceeds 0.05.

**

Phantom texture is unrealistic relative to clinical data.