Table 3.

Mass spectrometric (MS) parameters used for the LC-MS3 analysis of TMT-labeled peptides

Parameter Value
Polarity Positive
Source ESI
Spray Voltage 1800 V
Time between master scans (s) 3
Full MS
 Detector type Orbitrap
 Orbitrap Resolution 120000
 Automated gain control target (AGC Target) 2 × 105
 Maximum injection time (ms) 120
 Scan Range (m/z) 380–1500
dd-MS2
 Detector type Ion trap
 Ion trap speed Rapid
 Automated gain control target (AGC Target) 1 × 104
 Maximum injection time 50 ms
 Scan Range (m/z) 400–1200
 Activation type Collision induced dissociation (CID)
 CID energy(%) 35
 Monoisotopic peak determination Peptide
 Charge exclusion Include 2-6
 Dynamic precursor exclusion 30 s
dd-MS3
 Detector type Orbitrap
 Orbitrap Resolution 50000
 Automated gain control target (AGC Target) 5 × 104
 Maximum injection time (ms) 86
 Scan Range (m/z) 120–500
 Activation type Higher energy collision induced dissociation (CID)
 HCD energy (%) 65
 Precursor selection mode Synchronous precursor selection (SPS) Top 10