Table 4. Mechanical Performance of Medical Skin Adhesives Performed by an Instron 5568.

product yield stress (MPa) stress at break (MPa) elastic modulus (MPa) strain at break (%)
Sil2   0.71 ± 0.07 0.33 ± 0.07 803 ± 67
Mepilexa        
Eurotek (whole dressing) 0.20 ± 0.01 0.25 ± 0.02 1.37 ± 0.06 472 ± 29
Eurotek (adhesive layer alone)   0.15 ± 0.01 1.36 ± 0.12 98 ± 5
Salts   6.09 ± 0.30 3.34 ± 0.19 610 ± 21
a

The Mepilex has a three-layer structure composed of a microporous/uniform polyester thermoset film and perforated adhesive silicones. Because of the layered and perforated structure, mechanical tests produced unreliable measurements, where reliable stress/strain results that represent physical behavior could not be determined.