TABLE 6.

Dose ratio between the modified calibration and TRS‐398 using a cylindrical chamber (modified/TRS‐398cyl), and a parallel‐plate chamber (modified/TRS‐398parallel‐plate), modified calibration and TG‐51 using a cylindrical chamber (modified/TG‐51cyl), and using a parallel‐plate chamber (modified/TG‐51parallel‐plate) in multicenter radiotherapy.

Center Energy (MeV) Modified/ TRS‐398cyl Modified/ TRS‐398parallel‐plate Modified/ TG‐51cyl Modified/ TG‐51parallel‐plate
1 6 1.016 1.006
8 1.018 1.005
10 0.999 1.007 1.007 0.986
12 1.000 1.002 1.009 0.985
18 1.000 1.004 1.011 0.991
2 4 1.004 1.003
6 1.000 0.999
8 0.981 0.981
10 1.001 0.993 1.006 0.991
15 1.001 0.990 1.008 0.989
18 1.002 0.994 1.011 0.993
3 6 1.019 1.019
9 1.005 1.001
12 0.998 0.998 1.008 0.995
4 6 1.019 1.009
9 1.003 1.004
12 0.980 1.001 1.001 1.002
16 1.001 0.999 1.001 0.997
20 0.991 0.995 0.991 0.989
5 6 1.003 1.000
9 0.986 0.982
12 1.000 0.996 1.017 0.995
18 1.014 1.016 1.014 1.012
22 0.984 1.008 1.006 1.006
6 6 0.992 0.990
9 1.007 1.004
12 1.005 0.997 1.017 0.995
15 1.003 1.000 1.003 0.997
18 1.003 0.99 1.008 0.996
7 4 0.996 1.000
6 0.994 1.002
8 0.989 1.007
10 0.987 0.986 1.003 1.012
12 0.992 0.985 1.005 1.007
15 1.009 0.995 1.007 0.995