TABLE 7.

Estimated uncertainty budgets of absorbed dose to water that determined at maximum depth (zmax) under reference conditions using modified calibration method.

Component Uncertainty (%)
SSD setting 0.10
Positioning the chamber 0.17
Field size setting 0.10
PTP correction 0.10
Pioncorrection 0.10
Ppolcorrection 0.05
Peleccorrection 0.07
Humidity 0.05
Charge measurement 0.23
Beam quality conversion factor (kQ,Qo) 0.40
Total uncertainty 0.54