TABLE II:
EIT System | ACT5 | GE [26] | ACT 3 [8] | PulmoVista [16] | LuMon [15] | Enlight [17] |
---|---|---|---|---|---|---|
| ||||||
# of Electrodes | 16 / 32 | 16 / 32 | 32 | 16 | 32 | 32 |
Drive Pattern | Parallel | Parallel | Parallel | Pair | Pair | Pair |
Frame Rate (frames/sec) | 54 / 27 | 52 / 20 | 18.5 | 10–50 | 50 | 50 |
Signal Frequency (kHz) | 5–500 | 10 | 28 | 80–130 | 180–220 | 125 |
SNR (dB) | >100 | 110 | 90 | - | - | 50–95 |
Max. Current (rms) (1 electrode) (mA) | 0.35 | 0.11 | 0.5 | 9 | 0.7 – 3.7 | - |