TABLE 2.

Real-time PCR intra- and interassay variabilitya

Input copy no. per ml of blood Observed copy no.
Intra-assay (n = 4)b Interassay (n = 4)c
0 0 0
494 458 ± 22 (5) 521 ± 118 (23)
1,482 1,481 ± 426 (29) 1,445 ± 398 (28)
4,444 4,495 ± 1,226 (27) 4,914 ± 1,032 (21)
13,333 12,128 ± 4,594 (38) 11,776 ± 1,928 (16)
40,000 40,549 ± 3,462 (9) 44,309 ± 8,433 (19)
a

A standard panel of DBS with the numbers of 8E5 cells from 0 to 40,000 copies per ml of HIV-1-negative blood (3× serial dilutions) was prepared and extracted as described in Materials and Methods. For the intra-assay determination, DNAs from 24 samples (4 panels each with 6 spots with different 8E5 content) were extracted and then run in the same real-time assay. For the interassay determination, DNA samples from the panel were isolated in four separate extraction runs and then assayed in four separate real-time assays. Results are given as means±one standard deviation; coefficients of variation (%) are given in parentheses.

b

Proviral DNA number was determined by the dCT standard curve as described in the legend for Fig. 2C.

c

dCT-based standard curve was constructed for each interassay run.